Showing posts with label intro. Show all posts
Showing posts with label intro. Show all posts

24 May 2012

IMAT/VMAT basics

One of the newest and most interesting external beam delivery techniques today goes by many names: intensity modulated arc therapy (IMAT), volumetric modulated arc therapy (VMAT), etc. To make things more confusing, vendors have each given this technique their own proprietary names: RapidArc (Varian), SmartArc (Philips), and VMATTM [didn't I already say VMAT?] (Elekta). Maybe the most general and correct name for IMAT would be cone-beam dynamic angle fluence modulated x-ray therapy (CBDAFMXT), but you might confuse that acronym with a chemotherapy drug name... In this post I'll discuss some of the basics of this arc-based form of IMRT (and just call it IMAT to keep things simple).

At its most basic IMAT is essentially conventional IMRT, but with the gantry moving in one or more rotating arcs, rather than delivering from a small number of fixed angles. This means that most of the concepts and advantages and disadvantages of IMRT apply to IMAT (detailed below). IMAT was developed (and marketed!) as a conventional linac-based alternative to helical tomotherapy and as a more conformal / lower critical structure dose and faster version of static angle IMRT.

The hierarchy of IMRT techniques.
In the figure showing the IMRT hierarchy, IMAT is on the branch of cone-beam, dynamic gantry IMRT. In order to deliver IMAT, a linac must have some of the following capabilities: gantry motion with beam on, dynamic MLC (i.e. leaf motion with beam on and gantry rotating), and variable dose rate.

Planning of IMAT is very similar to conventional IMRT. The plan is determined by inverse planning methods. The degrees of freedom are increased by considering gantry rotation speed, dose rate, number of field shapes, number of arcs, etc. For planning, arcs are usually approximated with a finite number of angles (e.g. 36). Constraints can be more tightly matched with multiple arcs at the expense of delivery time. Another important aspect in IMAT optimization is that MLC leaf speed limits the beam shape "distance" from one angle to the next, i.e. the MLC leaf positions cannot vary greatly from one angle to the next and thus beam shape "interconnectedness" must be taken in to account.

Advantages of IMAT include:
  • Highly conformal target volume dose with lower dose to critical structures than IMRT or 3DCRT, as dose is spread over more angles.
  • Faster delivery times and lower MU's (especially single arc IMAT) when compared with IMRT.
  • Non-co-planar arcs possible.
  • Comparable plans to helical tomotherapy, but performed with a conventional linac.

Disadvantages of IMAT include:
  • Higher cost of hardware and software licensing relative to IMRT.
  • Increased complexity of plans makes QA a poor diagnostic tool (i.e. hard to determine source of QA failures).
IMAT delivery techniques are the obvious(?) next step following IMRT. In fact, it's hard to come up with a list of concrete disadvantages of IMAT over IMRT. (Please comment if you feel otherwise.) In our clinic it's one of the few new techniques that everyone seems to have adopted with open arms.

Further reading:

02 May 2012

The many faces of bolus

Bolus is a simple, yet important technology used in radiation therapy. The most basic function of bolus material is to shape the dose distribution in a desired way. This generally falls into two categories: compensating for "missing" tissue and enhancing the build-up effect of MeV energy photon beams.

The bolus itself can be made of a huge variety of materials depending on the application. Below are some materials used as bolus, all of which are applied directly to the skin surface.

"Superflab" vinyl plastic bolus in 5mm and 3mm thicknesses.

Edge view of Superflab sheets.

Wet towels can provide bolus with tissue like properties.

Petroleum jelly infused gauze pads are sometimes used as bolus used over open wounds.
Brass chainmail bolus for photons. The bolus of choice for C-3PO and Michael Jackson.

What's the strangest material you've seen used as bolus?

30 April 2012

Multileaf collimators: modern beam shaping

One of the most important aspects of therapeutic radiation delivery is beam shaping. The most common technology currently used to shape x-ray beams is the multileaf collimator or MLC.

120 leaf multileaf collimator.
MLC's consist of dozens of independently moving slats, or leaves, which can collimate the beam into nearly arbitrary shapes. MLC's were first introduced in the 1960's in Japan as a means of replacing conventional blocks in 3D conformal radiation therapy (3DCRT). Today they are standard on most therapy linacs and enable modern techniques based on intensity modulation.

Key advantages of MLC's include:

  • Dynamic movement of leaves during delivery allows for intensity modulated fluence patterns not achievable with conventional blocks.
  • Finite set of possible aperture shapes results in a reasonable solution space for inverse planning optimization for IMRT, VMAT, etc.
  • Significantly more convenient than cutting custom blocks for every field.

Drawbacks of MLC's include:
  • Large number of leaves increases possibility of mechanical failure (i.e. reliability issues).
  • Non-zero inter-leaf leakage.
  • Smoothness of shaping dependent on size of leaves and speed of leaf motors.

Almost everyone can agree that MLC's have been a huge advance in radiation therapy, with the advantages far outweighing the disadvantages. I plan to discuss MLC alternatives in a future post.


Image courtesy of Varian Medical Systems, Inc. All rights reserved. (source)

29 April 2012

The basics of IMRT: an overview

Intensity modulated radiation therapy (IMRT) is one of the workhorse delivery methods in current radiation therapy. In many ways, it is a significant step up over the previous standard technique of 3D conformal radiation therapy (3DCRT). I'm going to start talking about IMRT with a very basic overview of how it works.



IMRT is a technique to plan and deliver MeV range x-ray therapy. The main advance with IMRT over 3DCRT is the algorithmic optimization of dose from all delivery angles at the same time to meet a pre-defined set of objectives, so-called inverse planning. This is accomplished by using multileaf collimators, which can create arbitrary aperture shapes (within reason), thus modulating the dose to the targets. In turn, this gives the dose optimization algorithm a large number of parameters to work with to achieve the desired dose shape.


Key steps in IMRT planning and delivery:

  • Acquisition of patient geometry (via CT, MRI, etc).
  • Delineation of targets and avoidance volumes.
  • Beam angle and energy selection.
  • Optimization of fluences to desired prescription and avoidance objectives.
  • Assessment of DVH's.
  • QA via secondary calculations and field measurements.
Diagram of a multileaf collimator.

Further reading:

  • Bortfeld, IMRT: a review and preview, Phys. Med. Biol. 51 R363, 2006 doi:10.1088/0031-9155/51/13/R21
  • KY Cheung, Intensity modulated radiotherapy: advantages, limitations and future developments, Biomed Imaging Interv J 2006;2(1):e19 (open access)

Top image by Taheri-Kadkhoda et al. Radiation Oncology 2008 3:4 doi:10.1186/1748-717X-3-4 licensed under CC licensing.
Bottom image by ZEEs and licensed under CC licensing.

27 April 2012

365 Days of Medical Physics blogging

Welcome to my new blog! I'm intending to make short posts (nearly) every day for the next year covering a broad range of medical physics topics.

The idea behind this blog is largely to help motivate me to consistently push my medical physics knowledge, but also to foster discussion online relevant to others learning medical physics.

I am a medical physicist involved in radiation therapy, so this blog will center on issues related to clinical radiation therapy physics, but I will also venture into some diagnostic and research territory. For topics more closely related to my research on particle therapy and computational medical physics, I will probably post on Will Work for Science, the blog I co-author will Herr Dr. Niels Bassler.

I hope you enjoy the posts. Please comment, tell me how awesome this is, correct me, and/or make topic suggestions!