Multileaf collimators: modern beam shaping

One of the most important aspects of therapeutic radiation delivery is beam shaping. The most common technology currently used to shape x-ray beams is the multileaf collimator or MLC.

120 leaf multileaf collimator.
MLC's consist of dozens of independently moving slats, or leaves, which can collimate the beam into nearly arbitrary shapes. MLC's were first introduced in the 1960's in Japan as a means of replacing conventional blocks in 3D conformal radiation therapy (3DCRT). Today they are standard on most therapy linacs and enable modern techniques based on intensity modulation.

Key advantages of MLC's include:

  • Dynamic movement of leaves during delivery allows for intensity modulated fluence patterns not achievable with conventional blocks.
  • Finite set of possible aperture shapes results in a reasonable solution space for inverse planning optimization for IMRT, VMAT, etc.
  • Significantly more convenient than cutting custom blocks for every field.

Drawbacks of MLC's include:
  • Large number of leaves increases possibility of mechanical failure (i.e. reliability issues).
  • Non-zero inter-leaf leakage.
  • Smoothness of shaping dependent on size of leaves and speed of leaf motors.

Almost everyone can agree that MLC's have been a huge advance in radiation therapy, with the advantages far outweighing the disadvantages. I plan to discuss MLC alternatives in a future post.


Image courtesy of Varian Medical Systems, Inc. All rights reserved. (source)

29 April 2012

The basics of IMRT: an overview

Intensity modulated radiation therapy (IMRT) is one of the workhorse delivery methods in current radiation therapy. In many ways, it is a significant step up over the previous standard technique of 3D conformal radiation therapy (3DCRT). I'm going to start talking about IMRT with a very basic overview of how it works.



IMRT is a technique to plan and deliver MeV range x-ray therapy. The main advance with IMRT over 3DCRT is the algorithmic optimization of dose from all delivery angles at the same time to meet a pre-defined set of objectives, so-called inverse planning. This is accomplished by using multileaf collimators, which can create arbitrary aperture shapes (within reason), thus modulating the dose to the targets. In turn, this gives the dose optimization algorithm a large number of parameters to work with to achieve the desired dose shape.


Key steps in IMRT planning and delivery:

  • Acquisition of patient geometry (via CT, MRI, etc).
  • Delineation of targets and avoidance volumes.
  • Beam angle and energy selection.
  • Optimization of fluences to desired prescription and avoidance objectives.
  • Assessment of DVH's.
  • QA via secondary calculations and field measurements.
Diagram of a multileaf collimator.

Further reading:

  • Bortfeld, IMRT: a review and preview, Phys. Med. Biol. 51 R363, 2006 doi:10.1088/0031-9155/51/13/R21
  • KY Cheung, Intensity modulated radiotherapy: advantages, limitations and future developments, Biomed Imaging Interv J 2006;2(1):e19 (open access)

Top image by Taheri-Kadkhoda et al. Radiation Oncology 2008 3:4 doi:10.1186/1748-717X-3-4 licensed under CC licensing.